Date |
Hour |
ID |
Note |
Wed 01/Mar/06 |
11:00 |
GG |
After a long RECOVERING, Locked loops 1, 3-8. Pressure in these loops ~ 0.39 b. Weight 23.6 kg. |
11:37 |
IAC |
RUN. Standard procedure. |
12:45 |
SC |
Recovery |
17:00 |
GG |
Dummy staves switched OFF. Heaters damaged . System in STOP. No damages to the cooling lines. |
17:30 |
GG |
OFF loops 1, 3-8. System in RECOVERING. |
Thu 02/Mar/06 |
9:30 |
GG |
Weight 23.54. LOCKED loops 1, 2-10 @ ~0.39 b. |
10:28 |
GG |
RUN. |
10:36 |
GG |
Sector 0 ON , 12/12 HS |
10:40 |
GG |
Weight 18.25 kg, stable. |
11:15 |
GG |
Heater checked. All the 11 resistance still in specification (~5 Ohm, see paper logbook for details). Serious deformation of the carbon fibre support. |
16:16 |
GG |
Weight 18.15. Pressure in loop 1 0.74 b. Other still ~0.39 b. |
19:23 |
SC |
Recovery |
Fri 03/Mar/06 |
09:45 |
GG |
Weight 23.36 kg. Pressure in loop 1 0.66 b. Leak in a valve? Other still ~0.39 b. |
09:58 |
GG |
Temperature tube loop 10: injection and vapours 21°C (measured with quickTEMP, infrared thermometer). Loop 2 locked and 10 open. RUN |
10:12 |
GG |
Weight 18.2kg. |
10:30 |
GG |
Weight 17.4kg. linear decreasing. Temperature not changed. |
10:36 |
GG |
Weight 17.3kg. Temperature vapours 20°C. |
12:19 |
GG |
Weight 15.3kg. Temperature vapours 20°C. In ALARM for low volume in the condenser. RECOVERING |
Tue 07/Mar/06 |
08:00 |
GG |
General power cut. |
Thu 09/Mar/06 |
18:16 |
GG |
Starting up the system after the power cut. Chiller restarted, PVSS rebooted, System reset. Loop 4 ON. RECOVERING. |
Fri 10/Mar/06 |
15:00 |
GG |
LEAK TEST. |
Mon 13/Mar/06 |
09:52 |
GG |
Leak rates all high. Opened all the loops and RECOVERING. Weight 22.9 kg. |
Tue 14/Mar/06 |
16:54 |
GG |
LEAK TEST. |
17:12 |
GG |
The loops are NOT locked during the test, but closed. SW bug? |
17:26 |
GG |
Leak rates of ~ 1.8 mb/min. Weight 21.9 kg. Loops locked. STAND-BY. Pressure in the loops ~0.37 b |
Wed 15/Mar/06 |
15:08 |
GG |
System in RUN during the morning, up to now. |
15:09 |
GGRT |
Loop 2 closed, loop 10 open. RUN. Screwed the pressure sensor and the vacuum valves of pump 8a. |
15:50 |
GGRT |
Loop 2 open. RECOVERING. |
17:00 |
GGRT |
Refilling started during the recovering. Cylinder put at the top of a footboard. |
17:54 |
GGRT |
STOP. Refilling in progress. Acceleration of the refilling rate. |
19:00 |
GGRT |
Refilling completed at 32.6 kg. |
Fri 24/Mar/06 |
11:53 |
GG |
System in ALARM for a general power cut. Restored the compressor driver. Weight 32.5kg. Closed all loops. System in STOP. |
Wed 29/Mar/06 |
12:00 |
GG |
System found OFF for a local power cut. Restored. Weight 32.5 kg. Locked all loops. System in RECOVERING. |
15:19 |
SC |
System put in Run mode and after open loop2 |
15:39 |
SC |
System put in RECOVERING |
16:06 |
SC |
System put in Run Mode |
18:59 |
SC |
System put in RECOVERING |
Thu 30/Mar/06 |
09:31 |
GG |
Weight 30.3 kg. ~ 0.5 b in the closed loops. All loops open. |
14:54 |
SC |
Close all loops, put the system in run mode, then open the loop2 |
15:40 |
SC |
System in Recovering |
19:11 |
SC |
Allarm: "Tank Leak", aknoledge the allarm. Then system in Recovering |
Fri 31/Mar/06 |
10:20 |
SC |
System put in Run Mode and then open loop2 |
11:49 |
SC |
System put in Recovering |