-- PankajKharkwal - 2015-07-17

Scintillator Efficiency

Efficiency Measurement of Scintillators and PMT assembly

Experiment Setup: <br/>

Counter set to 5 Minutes


*PMT 17107*

Threshold of 40 mV

| High Voltage Applied (V) | And of all 3 PMT | And of Top & Bottom PMT | Efficiency | Standard Deviation (σ) |
| 500 | 1 | 481 | 0.20790021 | 0.00207684 |
| 525 | 2 | 462 | 0.43290043 | 0.00305444 |
| 600 | 137 | 487 | 28.1314168 | 0.02037515 |
| 625 | 318 | 478 | 66.5271967 | 0.02158399 |
| 650 | 410 | 459 | 89.3246197 | 0.01441355 |
| 700 | 450 | 472 | 95.3389831 | 0.00970297 |
| 750 | 463 | 499 | 92.7855711 | |
| 800 | 491 | 529 | 92.8166352 ||
| 850 | 454| 484 | 93.8016529 ||
| 900 | 449| 477 | 94.129979 ||
| 950 | 454| 478 | 94.9790795 ||
| 1000 | 437| 462 | 94.5887446 ||

Edit | Attach | Watch | Print version | History: r2 < r1 | Backlinks | Raw View | WYSIWYG | More topic actions